Device Recall Bionime Rightest Blood Glucose Monitoring System 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Invictus Scientific Inc 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    45953
  • 사례 위험등급
    Class 3
  • 사례 연번
    Z-0728-2008
  • 사례 시작날짜
    2007-11-26
  • 사례 출판 날짜
    2008-02-16
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2011-10-13
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Blood Glucose System - Product Code NBW
  • 원인
    Abnormal memory recall: when the meter is fixed in 12 hour mode, the time frame display of all memory data will be shown incorrectly as am or pm depending on when the user is recalling the memory stored in the meter. this will cause confusion as to when the last test was taken.
  • 조치
    An Urgent: Device Recall Notification letter were sent to customers (direct accounts) beginning on November 20, 2007, via US Postal Service, with a return receipt requested. The distributors were instructed to inform consumers/users to fix the time format to the 24 hrs mode and to contact the distributors or the firm for replacement, if they prefer to use the time format in the 12 hrs mode. They were also instructed that if they have further distributed any of the recalled products, to please immediately contact those accounts and advise them of the recall situation, and have them return their outstanding recalled stocks. If customers have any questions, they were instructed to contact at 858-481-8485.

Device

  • 모델명 / 제조번호(시리얼번호)
    Lot Numbers: 1FF19C0001 to 1FF19C0020,  1FF05C0001 to 1FF05C5000,  2FH15C0001 to 2FH15C4800,  230GBB0001 to 230GBB0050, 230GFF0001 to 230GFF0600, 230GFF0601 to 230GFF1200, 230GFF1201 to 230GFF1996, 230GHC0001 to 230GHC3204, 230GHC3205 to 230GHC4204, 230GHJ0001 to 230GHJ5000, 230GHJ5001 to 230GHJ6992, and 230GIB0001 to 230GIB3008
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Nationwide Distribution
  • 제품 설명
    Invictus-- Bionime Rightest Blood Glucose Monitoring System || MODEL: GM300; Invictus Scientific, Inc., San Diego, CA
  • Manufacturer

Manufacturer

  • 제조사 주소
    Invictus Scientific Inc, 11760 Sorrento Valley Rd, Ste G, San Diego CA 92121
  • Source
    USFDA