Sites affected will be those:
1. Running XiO Release 4.51 and higher, and
2. Using Elekta linacs, and
3. Using third party fixed wedges in combination with MLCs
Worldwide Distribution - USA (nationwide) and the country of Canada
제품 설명
XiO Radiation Treatment Planning System, Versions 4.30.00-4.64.00. || Used to create treatment plans for any cancer patient for whom external beam radiation therapy or brachytherapy has been prescribed. The system will calculate and display, both on-screen and in hard-copy, either two- or three-dimensional radiation dose distributions within a patient for a given treatment plan set-up.
Worldwide distribution: USA (nationwide) and countries of: China, France, Germany, Japan, Mongolia, Netherlands, Nicaragua, Poland, Russia, South Korea, Switzerland, and Turkey.
제품 설명
XiO Radiation Treatment Planning System, XiO Release 4.0.0 through 4.40.00 || Used to create treatment plans for any cancer patient for who external beam radiation therapy or brachytherapy has been prescribed.
Worldwide distribution: USA (nationwide) including: IL, IN, MA, NJ and OK; and country of Germany.
제품 설명
XiO Radiation Treatment Planning System, XiO versions equal to or greater than 2.4.0 and less than 4.63.0 || The XiO RTP System is used to create treatment plans for any cancer patient for whom external beam radiation therapy or brachytherapy has been prescribed.