Pacemaker Without Cables NANOSTIM

  • 모델명 / 제조번호(시리얼번호)
    NANOSTIM S1DLCP, specific serials.
  • 제품 설명
    The NANOSTIM wireless pacemaker is indicated for: 1. Chronic atrial fibrillation with second-degree AV block or bifascicular branch block. 2. Inaugural rhythms with second or third degree AV block or branch block and a low level of physical activity or a low expected life expectancy. 3. Sinus bradycardia with pausasaislated or unexplained syncope with electrophysiological results. Uses: the wireless pacemaker NANOSTIM is a pulse generator designed for bradycardia stimulation that incorporates a battery and electrodes and must be implanted in the right ventricle. Since it is a cordless device, it does not need a connector, stimulation cable or pulse generator pocket. A non-retractable distal helix with a single twist fixes the device to the endocardium. In addition, three sutures in the outer part of the head of the device offer a secondary fixation. Detection and stimulation occur between a distal electrode near the helix and the external housing of the device. The proximal end of the device has a mechanism that allows coupling it to the decolocation and extraction catheters for repositioning and extraction. The device is communicated bidirectionally with the programmer by electrical signals that are processed between the electrodes of the implanted device and the skin electrodes attached to the patient's chest and connected to the programmer. Accordingly, the device transmits signals by circuits and electrodes previously designed for stimulation, with data encoded by pulses during the refractory period of the heart. The device detects the blood temperature of the right ventricle and increases the frequency of stimulation as the metabolic requirement increases.
  • Manufacturer
  • 1 Event

제조사

  • 제조사 모회사 (2017)
  • 제조사 의견
    “We are in constant communication with regulatory agencies and competent authorities worldwide which allows us to implement global recalls or in-country communication quickly and effectively,” Abbott, which now owns St. Jude Medical told ICIJ in a statement. In addition to sending global notices to physicians worldwide, we also make sure that product advisories are available online and classification of product recalls and product advisories are determined by global regulatory bodies which can impact the timing in any given country. MD companies follow varying regulations in different countries. In come countries software is not regulated so a recall in one country related to software would not be classified as a recall or field action in another. In addition, review cycles within the regulatory process can be different in each country which can impact communication and recall timing.
  • Source
    NIDFSINVIMA

유사한 이름을 가진 의료기기 1건 유사한 이름을 가진 의료기기 1 건

데이터에 대해 더 자세히 알아보기 여기

  • 모델명 / 제조번호(시리얼번호)
  • Manufacturer
    N/A