Ingenuity Core

  • 모델명 / 제조번호(시리얼번호)
  • 제품 설명
    Medical Device Safety Alert: Philips Ingenuity Core
  • Manufacturer
  • 1 Event

제조사

  • 제조사 모회사 (2017)
  • Source
    DH

유사한 이름을 가진 의료기기 1건 유사한 이름을 가진 의료기기 161 건

데이터에 대해 더 자세히 알아보기 여기

  • 모델명 / 제조번호(시리얼번호)
    Head Holder M/N 4535 671 11331. Ingenuity Core - M/N 728321, Serial #s: 7725, 10614, 10663, 50171, 95735, 300072, 310001, 310006, 310008, 310009, 310010, 310012, 310013, 310014, 310015, 310017, 310018, 310020, 310021, 310022, 310024, 310025, 310026, 310028, 310029, 310030, 310031, 310032, 310033, 310034, 310035, 310037, 310038, 310039, 310040, 310041, 310042, 310043, 310044, 310045, 310046, 310047, 310048, 310049, 310050, 310051, 310052, 310053, 310054, 310055, 310056, 310057, 310058, 310059, 310060, 310061, 310062, 310063, 310064, 310065, 310066, 310067, 310068, 310069, 310070, 310071, 310072, 310073, 310074, 310075, 310076, 310077, 310078, 310079, 310080, 3100016, EP16E110078, EP16E120020 & EP16ER110018.
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution-Product was shipped to the following states: AR, AR, AZ, CA, CO, CT, FL, GA, IA, IN, KS, LA, MA, MD, MI, MO, MS, NC, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA, WV & WY, and to the following countries: Argentina, Australia, Bangladesh, Belgium, Brazil, Bulgaria, China, Colombia, Denmark, Ecuador, France, Germany, India, Indonesia, Israel, Japan, Latvia, Libya, Myanmar, Norway, Oman, Poland, Republic of Korea, Russian Federation, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom & Viet Nam.
  • 제품 설명
    Head Holder shipped with Ingenuity Core Computed Tomography X-ray Systems. || Intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.
  • Manufacturer
  • 모델명 / 제조번호(시리얼번호)
    Model #: 728321; Serial #s: 52000, 52001, 52002, 52004, 52006, 52007, 52008, 52009, 52010, 52013, 52015, 52016, 52017, 52018, 52020, 52021, 52022, 52024, 52026, 52033, 52038, 52039, 52046, 52051, 52052, 52057, 310001, 310002, 310003, 310004, 310005, 310006, 310007, 310009, 310010, 310011, 310012, 310013, 310014, 310015, 310016, 310017, 310018, 310019, 310020, 310021, 310022, 310024, 310025, 310026, 310027, 310028, 310029, 310030, 310031, 310032, 310033, 310034, 310035, 310036, 310037, 310038, 310040, 310041, 310043, 310044, 310045, 310046, 310047, 310048, 310049, 310050, 310051, 310052, 310053, 310054, 310055, 310056, 310057, 310059, 310060, 310061, 310062, 310063, 310064, 310065, 310066, 310068, 310069, 310070, 310071, 310072, 310073, 310074, 310075, 310076, 310077, 310078, 310079, 310080, 310081, 310082, 310083, 310084, 310085, 310086, 310087, 310088, 310089, 310090, 310091, 310092, 310093, 310094, 310095, 310096, 310097, 310099, 310100, 310101, 310102, 310103, 310104, 310105, 310106, 310107, 310108, 310109, 310110, 310111, 310112, 310113, 310114, 310115, 310116, 310117, 310118, 310119, 310120, 310121, 310123, 310124, 310125, 310126, 310128, 310129, 310131, 310132, 310133, 310134, 310135, 310136, 310137, 310138, 310139, 310140, 310141, 310143, 310144, 310145, 310148, 310149, 310151, 310152, 310153, 310154, 310155, 310157, 310158, 310160, 310161, 310162, 310163, 310164, 310165, 310170, 310171, 333001, 333002, 333003, 333005, 333006, 333007, 333008, 333009, 333010, 333011, 333012, 333013, 333014, 333015, 333016, 333017, 333022, 333024, 333025, 333026, 333027, 333030, 333031, 333035, 333037, 333038, 333042, 333044, 333045, 333046, 333051, 333052, 333053, 333055, 333057, 333060, 333061, 333066.
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution-US (nationwide) including DC & Puerto Rico and the following countries: Albania, Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia, Brazil, Bulgaria, Cambodia, Cameroon, Cayman Islands, China, Colombia, Democratic Congo, Croatia, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Ethiopia, Finland, France, Gabon, Georgia, Germany, Ghana, Greece, Guatemala, Hong Kong, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Kuwait, Kyrgyzstan, LAO People's Dem. Rep., Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malawi, Malaysia, Malta, Mauritius, Monaco, Mongolia, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Philippines, Poland, Portugal, R¿union, Romania, Russia, Rwanda, Saudi Arabia, Senegal, Serbia & Montenegro, Singapore, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Vietnam, Yemen, and Zambia.
  • 제품 설명
    Ingenuity Core (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.
  • Manufacturer
  • 모델명 / 제조번호(시리얼번호)
    Model #: 728323; Serial #s: 30153, 32002, 32003, 32004, 32005, 32006, 32007, 32009, 32010, 32011, 32013, 32014, 32015, 32016, 32018, 32021, 32023, 32025, 32026, 32029, 32030, 32031, 32032, 32033, 32034, 32035, 32036, 32037, 32038, 32040, 32041, 32042, 32043, 32044, 32046, 32047, 32048, 32049, 32051, 32052, 32053, 32054, 32057, 32059, 32060, 32062, 32064, 32066, 32068, 32069, 32070, 32072, 32073, 32074, 32077, 32078, 32080, 32083, 32084, 32089, 32091, 32092, 32094, 32095, 32098, 32100, 32101, 32104, 32106, 32107, 32108, 300137, 310067, 320001, 320002, 320003, 320004, 320005, 320006, 320007, 320008, 320009, 320010, 320011, 320013, 320014, 320015, 320018, 320019, 320020, 320021, 320022, 320023, 320024, 320025, 320026, 320027, 320028, 320029, 320030, 320031, 320032, 320033, 320034, 320035, 320036, 320037, 320038, 320039, 320040, 320041, 320043, 320044, 320045, 320046, 320047, 320048, 320049, 320050, 320051, 320052, 320053, 320054, 320055, 320056, 320057, 320059, 320060, 320062, 320063, 320064, 320065, 320066, 320067, 320068, 320069, 320070, 320071, 320072, 320073, 320074, 320075, 320076, 320077, 320078, 320079, 320080, 320081, 320082, 320083, 320084, 320085, 320086, 320087, 320088, 320089, 320090, 320092, 320093, 320095, 320096, 320097, 320098, 320099, 320101, 320102, 320103, 320104, 320105, 320106, 320107, 320109, 320110, 320111, 320112, 320114, 320115, 320116, 320117, 320118, 320119, 320120, 320121, 320122, 320123, 320124, 320125, 320126, 320128, 320129, 320131, 320132, 320133, 320134, 320135, 320137, 320138, 320139, 320140, 320141, 320142, 320143, 320144, 320145, 320146, 320147, 320148, 320150, 336001, 336002, 336003, 336006, 336007, 336008, 336009, 336010, 336011, 336012, 336013, 336014, 336015, 336016, 336018, 336020, 336021, 336024, 336027, 336028, 336029, 336030, 336031, 336032, 336033, 336034, 336035, 336036, 336037, 336038, 336039, 336040, 336042, 336043, 336044, 336046, 336047, 336048, 336050, 336052, 336055, 336056, 336057, 336058, 336059, 336061, 336063, 336066, 336067, 336068, 336069, 336070, 336071, 336072, 336074, 336075, 336076, 336080, 336081, 336086, 336087, 336089, 336702, 336703, 336704, 336705, 336706, 336708, 336709, 336710, 336711, 336713, 336714, 336715, 336716, 336717, 336718, 336719, 336721, 336707
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution-US (nationwide) including DC & Puerto Rico and the following countries: Albania, Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia, Brazil, Bulgaria, Cambodia, Cameroon, Cayman Islands, China, Colombia, Democratic Congo, Croatia, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Ethiopia, Finland, France, Gabon, Georgia, Germany, Ghana, Greece, Guatemala, Hong Kong, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Kuwait, Kyrgyzstan, LAO People's Dem. Rep., Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malawi, Malaysia, Malta, Mauritius, Monaco, Mongolia, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Philippines, Poland, Portugal, R¿union, Romania, Russia, Rwanda, Saudi Arabia, Senegal, Serbia & Montenegro, Singapore, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Vietnam, Yemen, and Zambia.
  • 제품 설명
    Ingenuity Core 128 (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.
  • Manufacturer
  • 모델명 / 제조번호(시리얼번호)
    Head Holder M/N 4535 671 11331. Ingenuity Core 128 - M/N 728323, Serial #s: 300068, 320003, 320004, 320005, 320006, 320007, 320008, 320009, 320010, 320011, 320013, 320014, 320015, 320017, 320018, 320019, 320020, 320021, 320022, 320023, 320024, 320025, 320026, 320027, 320028, 320029, 320030, 320031, 320032, 320033, 320034, 320035, 320036, 320037, 320038, 320040, 320041, 320042, 320043, 320044, 320045, 320046, 320047, 320048, 320049, 320050, 320052, 320053, 320054, 320055, 320056, 320057, 320058, 320059 & EP16E110092.
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution-Product was shipped to the following states: AR, AR, AZ, CA, CO, CT, FL, GA, IA, IN, KS, LA, MA, MD, MI, MO, MS, NC, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA, WV & WY, and to the following countries: Argentina, Australia, Bangladesh, Belgium, Brazil, Bulgaria, China, Colombia, Denmark, Ecuador, France, Germany, India, Indonesia, Israel, Japan, Latvia, Libya, Myanmar, Norway, Oman, Poland, Republic of Korea, Russian Federation, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom & Viet Nam.
  • 제품 설명
    Head Holder shipped with Ingenuity Core 128 Computed Tomography X-ray Systems. || Intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.
  • Manufacturer
  • 모델명 / 제조번호(시리얼번호)
    Model #728321; Serial #s: 31001, 52007, 52008, 52010, 52012, 52013, 52018, 52020, 52021, 52024, 52026, 52030, 52031, 52032, 52033, 52034, 52035, 52036, 52037, 52038, 52039, 52042, 52043, 52046, 52052, 52053, 52054, 52055, 52057, 52058, 310001, 310002, 310003, 310009, 310017, 310052, 310055, 310059, 310064, 310069, 310070, 310071, 310081, 310087, 310088, 310096, 310111, 310113, 310119, 310121, 310129, 310133, 310134, 310141, 310143, 310144, 310149, 310151, 310157, 310160, 310170, 333002, 333009, 333014, 333015, 333030, 333031, 333037, 333042, 333043, 333044, 333045, 333046, 333051, 333052, 333053, 333055, 333056, 333057, 333059, 333060 & 333061.
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution-US (nationwide) including the states of AR, AZ, CA, CT, DE, FL, GA, IA, IN, KS, LA, MA, MD, MI, MN, MO, MS, MT, NH, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, VT, WV & WY, and to the following countries: Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Cambodia, Canada, China, Colombia, Denmark, Finland, France, Germany, Hong Kong, India, Indonesia, Iran, Iraq, Ireland, Italy, Japan, Latvia, Libyan Arab Jamahiriya, Lithuania, Malaysia, Mexico, Myanmar, Netherlands, New Zealand, Norway, Panama, Philippines, Russia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uzbekistan & Viet Nam.
  • 제품 설명
    Computed Tomography X-ray Systems Ingenuity Core. || Intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.
  • Manufacturer
하나 더 156 건 더