Device Recall Spacelabs Healthcare Pathfinder SL Holter Analyzer

  • 모델명 / 제조번호(시리얼번호)
    Version 1.6.0.
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution - US (nationwide) including the states of Nevada and Vermont, and the countries of AUSTRALIA, AUSTRIA, CANADA, CYPRUS, GERMANY, ICELAND, ISRAEL, POLAND, ROMANIA, SOUTH AFRICA, SWITZERLAND, THAILAND, TURKEY and UNITED KINGDOM.
  • 제품 설명
    The product is Pathfinder SL, version 1.6.0.3057 software used with the Spacelabs Healthcare Pathfinder SL Holter Analyzer, version 1.6.0. || Product Usage: || The Holter Analyzer analyzes information collected from ambulatory electrocardiogram recorders worn by patients and used to assist physicians in diagnosis and patient monitoring. In addition, incorrect QTc values are reported after moving the cursor. QTc values are timing events between labeled points of interest in the ECG signal that the physician can mark with the cursor. The software is used to edit a family of events and upon conclusion the software crashes with an error message.
  • Manufacturer
  • 1 Event

제조사

  • 제조사 주소
    Spacelabs Healthcare, Llc, 5150 220th Ave Se, Issaquah WA 98029-6834
  • 제조사 모회사 (2017)
  • Source
    USFDA