Device Recall RestoreUltra and RestoreSensor

  • 모델명 / 제조번호(시리얼번호)
    The first released affected product was released to the market in April, 2005.
  • 의료기기 분류등급
  • 의료기기 등급
    3
  • 이식된 장치?
    Yes
  • 유통
    Nationwide Distribution - all states including DC and Puerto Rico.
  • 제품 설명
    Medtronic neurostimulators. Restore (37711), RestoreADVANCED (37713), RestoreADVANCED SureScan MRI (97713), RestoreSENSOR (37714), RestoreSENSOR SureScan MRI (97714), PrimeAdvanced SureScan MRI, (97702), RestoreULTRA (37712), RestoreULTRA SureScan MRI (97712), Itrel 4 (37703), Itrel 4 (37704), RestorePrime (37701), PrimeAdvanced (37702), Spinal Cord Stimulators for chronic pain.
  • Manufacturer
  • 1 Event

제조사

  • 제조사 주소
    Medtronic Neuromodulation, 7000 Central Ave NE, Minneapolis MN 55432-3568
  • 제조사 모회사 (2017)
  • 제조사 의견
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.
  • Source
    USFDA