Device Recall PercuNav MP008RB Button Probe

  • 모델명 / 제조번호(시리얼번호)
    Lot codes and Expiry Date: I280903 (09-2012); C151005 (03-2013); E111003 (05-2013); and E251002 (05-2013).
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Nationwide Distribution: in the states of FL, IL, MA, NE, NY, SC, and VA. There are no foreign consignees and US government consignees.
  • 제품 설명
    The PercuNav MP008R-B Button Probe is designed to be used with Philips (formerly Traxtal) PercuNav Image Guided Intervention System. Some of the clinical applications of the Button Probe require sterility and others may not. However, it is possible that patient contamination could occur if a non-sterile Button Probe is used in clinical applications where aseptic technique is required. || The device's labeled in part: " TRAXTAL PERCUNAV...(1) Button Probe Right hand version, 25mm...Traxtal Inc...49 Spadina Avenue, Suite 310 Toronto, ON Canada M5V 2J1...Made In Canada...www.traxtal.com...STERILE...For use with Traxtal's PercuNav system only. Do not use if package is opened or damaged. Dispose of safely after single use to avoid risk of infection...Rx only...Use immediately after opening unit package. Store at room temperature".
  • Manufacturer
  • 1 Event

제조사

  • 제조사 주소
    Philips Healthcare Inc., 3000 Minuteman Road, Andover MA 01810
  • 제조사 모회사 (2017)
  • Source
    USFDA