Ingenuity CT Computed Tomography Xray system

  • 모델명 / 제조번호(시리얼번호)
    30003, 30004, 30008, 30014, 30033, 30039, 30040, 30041, 30042, 30043, 30044, 300037, 300066, 300097, 300151, 300156, 300157, 300203, 300204, 300205, 300206, 300207, 300208, 300209, 300210, 300211, 300212
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Worldwide distribution. The products were shipped to the following states: AL, AZ, CO, GA, IN, KY, LA, MA, MN, NY, OH, PA, SC, TX, and UT. There are five (5) government accounts for this recall: DEPARTMENT OF VETERANS AF, VETERANS WAY AND LAMONT ST, Mountain Home, TN 37684; US ARMY EVANS COMMUNITY H, 1650 Cochrane Cir, Colorado Springs, CO 80913; US VETERANS ADMINISTRATION, 3710 SW US VETERANS HOSPITAL, PORTLAND, OR 97239; VA MEDICAL CENTER, 4100 W 3RD ST, DAYTON, OH 45428; VAMC - PHILADELPHIA, UNIVERSITY & WOODLAND AVE, PHILADELPHIA, PA 19104. There are five (5) Canadian accounts for this recall: FIVE HILLS HEALTH REGION, 55 DIEFENBAKER DR, MOOSE JAW, SASKATCHEWAN S6J 0C2; INSTITUT UNIV CARDIOLOGIE, 2725 CH STE FOY, SAINTE FOY, QUEBEC G1V 4G5; LAKERIDGE HEALTH, 1 HOSPITAL CRT, OSHAWA, ONTARIO L1G 2B9; LAKERIDGE HEALTH BOWMANVI, 47 LIBERTY ST S, BOWMANVILLE, ONTARIO L1C 2N4; MCGILL UNIVERSITY HEALTH, 1051 BOUL D¿CARIE, MONTR¿AL, QUEBEC H4A 0A5. There were zero (0) Mexican accounts for this recall. The recalled products were shipped to the following countries: Argentina, Australia, Belgium, China, Egypt, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Republic of Korea, Lebanon, Netherlands, New Zealand, Norway, Russian Federation, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, and United Kingdom.
  • 제품 설명
    Ingenuity CT Computed Tomography X-ray system, Model 728326
  • Manufacturer
  • 1 Event

제조사

  • 제조사 주소
    Philips Medical Systems (Cleveland) Inc, 595 Miner Rd, Cleveland OH 44143-2131
  • 제조사 모회사 (2017)
  • Source
    USFDA

유사한 이름을 가진 의료기기 1건 유사한 이름을 가진 의료기기 3 건

데이터에 대해 더 자세히 알아보기 여기

  • 모델명 / 제조번호(시리얼번호)
    Head Holder M/N 4535 671 11331. Ingenuity CT M/N 728326, Serial #s: 6617, 10607, 10689, 32000, 300043, 300064, 300068, 300071, 300072, 300073, 300074, 300075, 300076, 300077, 300078, 300079, 300080, 300081, 300082, 300084, 300085, 300086, 300087, 300088, 300089, 300091, 300092, 300093, 300094, 300095, 300096, 300097, 300098, 300099, 300100, 300101, 300102, 300103, 300104, 300105, 300106, 300107, 300108, 300109, 300110, 300111, 300112, 300113, 300114, 300116, 300117, 300118, 300119, 300120, 300121, 300122, 300123, 300124, 300125, 300126, 300127, 300128, 300129, 300130, 300131 & 300132.
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution-Product was shipped to the following states: AR, AR, AZ, CA, CO, CT, FL, GA, IA, IN, KS, LA, MA, MD, MI, MO, MS, NC, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA, WV & WY, and to the following countries: Argentina, Australia, Bangladesh, Belgium, Brazil, Bulgaria, China, Colombia, Denmark, Ecuador, France, Germany, India, Indonesia, Israel, Japan, Latvia, Libya, Myanmar, Norway, Oman, Poland, Republic of Korea, Russian Federation, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom & Viet Nam.
  • 제품 설명
    Head Holder shipped with Ingenuity CT Computed Tomography X-ray Systems
  • Manufacturer
  • 모델명 / 제조번호(시리얼번호)
    System Code #728231, 4003, 4021, 9033, 9037, 9041, 9045, 9048, 9062, 9064, 9070, 9075, 9080, 9108, 9120, 9142, 9151, 9196, 9197, 9201, 9202, 9205, 9206, 9208, 9214-9218, 9462, 9508, 9531, 9534, 9535, 9544, 9554, 9561, 9584, 9603, 9606, 9608, 9613, 9627, 9630, 9657, 9668, 9671, 9672, 9695, 9706, 9708, 9719, 9728, 9734, 9735, 9741, 9745, 9756, 9789, 9819, 9832, 9838, 9839, 9840, 9859, 9874, 9875, 9891, 9898, 9904, 9918, 9926, 9952, 9964, 9965, 9968, 9975, 9997, 9999, 10000, 10008, 10015, 10041, 10052, 10054, 10092, 10095, 10100, 10103, 10114, 10125, 10145, 10151, 10152, 10154, 10170-10172, 10192, 10196, 10198, 10221, 10222, 10251, 10256, 10273, 10286, 10294, 10307, 10312, 10336, 10337, 10350, 10352, 10362, 10363, 10372, 10376, 10380, 10386, 10393, 10412, 10425, 10426, 10428, 10438, 10445, 10446, 10452, 10459, 10461, 10468, 10470, 10471, 10473, 10482, 10486, 10489, 10491, 10493, 10497, 10503, 10505, 10511-105531, 10533-10555, 10557-10566, 10568-10575, 10577-10604, 10606, 10608-10620, 10622-10624, 10628-10640, 29037, 29049, 29057, 64040, 90036, 90054, 90058, 90063, 90080, 90100, 90102, 90106, 90117, 90135, 90169, 90182, 91003, 95015, 95055, 95065, 95074, 95077, 95101, 95108, 95130, 95133, 95138, 95140, 95157, 95168, 95174, 95178, 95184, 95185, 95191, 95207, 95243, 95295, 95315, 95324, 95327, 95350, 95357, 95358, 95362, 95389, 95403, 95414, 95425, 95429, 95443, 95455, 95504, 95516, 95529, 95534, 95568, 95570, 95572, 95573, 95575, 95602, 95605, 95609, 95613, 95624, 95630, 95638, 95652, 95659, 95664, 95665, 95677, 95684, 95687, 95697, 95700-95728 & 300074.  System Code #: 728326, Serial #: 9511, 300003-300005, 300010-300022, 300024, 300025, 300027-300066 & 300068-300073.
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution-USA (nationwide) including the states of AK, AR, AZ, CA, CO, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MT, ND, NE, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA and WV, and the countries of Australia, Bahamas, Bangladesh, Belgium, Bolivia, Brazil, Chile, China, Colombia, Croatia, Denmark, Finland, France, Georgia, Germany, Great Britain, India, Indonesia, Iraq, Israel, Italy, Japan, Kazakhstan, Korea, Laos, Lebanon, Lithuania, Malaysia, Mexico, Netherlands, New Zealand, Norway, Poland, Russia, Slovakia, South Africa, Spain, Sudan, Sweden, Switzerland, Syria, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom and Vietnam.
  • 제품 설명
    Brilliance 64 and Ingenuity CT Computed Tomography X-Ray Systems. Brilliance 64 Model# 728231, Ingenuity CT Model # 728326, Distributed By: Philips Healthcare, Cleveland, OH. || The Brilliance 64 and Ingenuity CT are Computed Tomography X-Ray Systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories.
  • Manufacturer
  • 모델명 / 제조번호(시리얼번호)
    Model No. 728326; S/N: 2, 1234, 30005, 30009, 30017, 30018, 30020, 30023, 30028, 30031, 30034, 30037, 31001, 32065, 32067, 52030, 52031, 52044, 52049, 52050, 300159, 300204, 300205, 300207, 300208, 300209, 310002, 310003, 333057, 333064 & 336061.
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Worldwide distribution. US nationwide, Australia, Austria, Bangladesh, Belgium, Brazil, Cambodia, Canada, China, Egypt, France, Germany, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Korea, Lebanon, Libya, Lithuania, Malaysia, Martinique, Myanmar, Mexico, Mozambique, Netherlands, New Zealand, Norway, Palestine, Panama, Philippines, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Switzerland, Tanzania, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uzbekistan & Vietnam.
  • 제품 설명
    Ingenuity CT Computed Tomography X-ray system
  • Manufacturer