12 Instrument Sterilisation Tray (used for the Conquest Manual Instrument line) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Stryker Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00480-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2014-04-29
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Stryker has become aware that gravity sterilisation parameters provided in the 12 instrument sterilisation reprocessing guide (1000400920 rev d) did not fully consider all worst-case scenarios during validation testing. although, to date there have been no reported events to stryker of infection as a result of the use of the sterilization parameters previously provided with these trays, there is a remote possibility for infection.
  • 조치
    End users are advised to discard any old Instructions for Use (IFU) for 12 Instrument Sterilization Tray(s) that make reference to the gravity steam sterilization method (121oC for 30 minutes). New IFU can be downloaded from www.stryker.com/12instrumenttray