2-Way Foley Catheters and 2-Way Council Tip Foley Catheters (used for urinary drainage)Distributed between June 2013 to June 2015 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Bard Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-01254-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2015-12-22
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Bmd has identified that the instructions for use (ifu) included with the affected products provide the incorrect inflation volume. the inflation volume provided on the catheter’s valve cap does state the correct inflation volume. the ifu contains a chart entitled recommended inflation volumes with incorrect volumes. the inflation volume provided on the catheters valve cap does state the correct inflation volume.If the user follows the ifu and under inflates the balloon, the patient may experience some form of urine bypass due to the poor seating of the balloon in the bladder.
  • 조치
    Bard is providing users with the correct inflation volumes for the affected products. This action has been closed-out on 30/08/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    2-Way Foley Catheters and 2-Way Council Tip Foley Catheters (used for urinary drainage)Distributed between June 2013 to June 2015Multiple product and catalogue numbers affectedARTG Number: 158417
  • Manufacturer

Manufacturer