4FR Single Lumen Pressure Injectable Peripherally Inserted Central Catheter (PICC) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Teleflex Medical Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00915-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2015-09-15
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    There is a labelling inaccuracy in the instructions for use (ifu) and injection card. the ifu and the injection card contained within the product states that picc testing included 10 pressure injection cycles. this is not correct, it should state that catheter testing included only 5 pressure injection cycles.If it were to be used for more than five (5) pressure injections, it cannot be ruled out that there is a potential risk that the catheter could rupture during injections 6-10. catheter rupture can cause catheter or air embolism, bleeding or ineffective treatment and may require the need for radiologic or surgical intervention. no customer complaints have been received for this issue.
  • 조치
    Customers are notified not to use the device for more than five pressure injections. Future stock will have the IFU and injection card amended to reflect this change. This action has been closed-out on 26/08/2016.