ABGII Modular Stems and ABGII Modular Necks (An exchangeable modular neck, designed to allow intra-operative flexibility for bio mechanical adjustment of offset and anteversion without affecting leg length) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Stryker Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2012-RN-00493-3
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2012-07-06
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    This action is occurring worldwide because of a high incidence of wear, known as fretting and corrosion, at the join between the femoral stem and the neck component. fretting and corrosion can result in metal debris in and around the hip joint. if this occurs, patients may experience inflammation, adverse local tissue reactions (altrs), hypersensitivity/allergic reactions and osteolysis (bone loss).
  • 조치
    The ABGII Modular Femoral Stem and Modular Neck has been cancelled from the Australian Register of Therapeutic Goods (ARTG). Stryker Australia Pty Ltd has contacted orthopaedic surgeons who has ever implanted the ABGII modular hip replacement system. Those surgeons have been asked to contact their patients to discuss what clinical action is most appropriate. For more details, please see https://www.tga.gov.au/alert/abgii-modular-femoral-stem-and-modular-neck-hip-replacement-system and https://www.tga.gov.au/alert/abgii-modular-femoral-stem-and-modular-neck-hip-replacement-system-update .

Device

  • 모델명 / 제조번호(시리얼번호)
    ABGII Modular Stems and ABGII Modular Necks (An exchangeable modular neck, designed to allow intra-operative flexibility for bio mechanical adjustment of offset and anteversion without affecting leg length) Multiple Catalogue NumbersAll lot numbersARTG Number: 145647
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • Source
    DHTGA