Access AMH QC Kits for use with Access Family of Immunoassay Systems. An in vitro diagnostic medical device (IVD) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Beckman Coulter Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00859-1
  • 사례 위험등급
    Class III
  • 사례 시작날짜
    2016-06-29
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Beckman coulter has identified that the quality control (qc) value card includes incorrect standard deviation values for the access amh qc kit, lot 689202. the qc value card for this lot lists the following standard deviation values: qc1 as 0.02, qc2 as 0.11 and qc3 as 0.65. the correct standard deviation values are as follows: qc1 is 0.15, qc2 is 0.76 and qc3 is 2.30. use of this restrictive qc range may result in qc failures but there is no impact to patient results.
  • 조치
    Beckman Coulter is advising users to use the correct QC values, as provided, when establishing QC ranges for the affected lot. Users should also confirm that the Access AMH QC assigned mean and standard deviation values are set appropriately per individual laboratory procedures. This action has been closed-out on 20/02/2017.

Device

  • 모델명 / 제조번호(시리얼번호)
    Access AMH QC Kits for use with Access Family of Immunoassay Systems. An in vitro diagnostic medical device (IVD).Reference Number: B13129Lot Number: 689202Expiry date: 16 September 2017ARTG Number: 213976
  • Manufacturer

Manufacturer