Accu-Chek Mobile Blood Glucose Meter 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Roche Diagnostics Australia Pty Limited 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00592-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2014-05-28
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Roche diagnostics has identified that a small number of people with diabetes using the accu-chek mobile device have experienced falsely elevated blood glucose readings when using the system as a result of not following the described and labelled handling instructions. such improper handling can include, for example, unclean hands contaminated with glucose-containing substances, pressing the finger too hard and too long on the test field or smearing the blood while performing a test. roche has decided to enhance its training and handling instructions for the accu-chek mobile system – increasing their visibility in the meter kit to ensure people with diabetes’ understanding of how to appropriately use the system. this will enable the users to obtain safe and reliable blood glucose (bg) readings at all times.Br /br /the tga has published a web statement regarding this issue. for more information, pleaser refer to: http://www.Tga.Gov.Au/safety/alerts-device-accu-chek-glucose-meter-140606.Htm.
  • 조치
    Consumers are advised to continue to use their Accu-Chek Mobile device safely and reliably by carefully following the instructions that are described in the User Manual and which can also be found in a new instructions leaflet available at www.accu-chek.com.au. Specifically: · Wash your hands with warm water and soap. Dry your hands thoroughly before obtaining a blood sample. · Form a proper blood drop and apply it to the centre of the test field. · Immediately apply the blood gently to the test field after you have created the blood drop. · Do not press against the test field on the tape. · Touch the test area gently and your finger should be removed from the test field when the beep tone sounds and/or “test in progress” is displayed.