Accu2ipMTA Applicator 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Getz Healthcare Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00759-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-06-21
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The supplier angiodynamics has identified that the accu2i pmta applicator has the potential to not function as expected during use. the desired microwave energy will not be delivered due to coolant ingress in an electrical connection, this condition results in an error code “high reflective power” displayed on the microwave generator. please note that not all “high reflective power” error codes are a result of this failure mode, as this error code results from any instance where microwave energy is not efficiently transmitted to the tissue. although the specific failure mode of the applicator not delivering the desired microwave energy would not result in any direct harm to the patient; the inserted applicator will need to be removed and replaced in order to continue the procedure.
  • 조치
    Customers are asked to inspect stocks, quarantine all affected units and complete the supplied Facsimile Reply Form to advise as to whether or not you still have stocks of the goods relevant to this recall.