ACCURUS Vitrectomy probes 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Alcon Laboratories Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00481-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2015-06-05
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Alcon has identified an insufficient seal on the outside packaging that could potentially affect the sterility of the product. the use of non-sterile vitrectomy probes in surgery has the potential to result in patient infection or inflammation and may lead to serious injury and sequelae, including loss of vision.. at this time, no adverse events or complaints have been confirmed related to the accurus vitrectomy probe lots impacted by this issue. regular post-operative patient follow-up may enable the surgeon to detect early potential abnormal inflammatory reaction and/or infection, which may reduce the severity of the ensuing event.
  • 조치
    Customers are advised to immediately stop further use of the affected lots and to segregate them to ensure that they are not used. Alcon will provide reimbursement for all returned product This action has been closed-out on 9/08/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    ACCURUS Vitrectomy probesACCURUS 23GA Standalone Vitreous ProbeCatalog # 8065750821Lot number: 14017912XAnterior ACCURUS® Probe with infusion needle Catalog # 8065803650Lot numbers: 14010430X, 140174S1X, 14010430X, 14017451XARTG Numbers 140947 and 145666