ACUSON SC2000 ultrasound with 5.0 software (VB20A, VB20B, or VB20C) and the ECG (physio) hardware and Auxiliary (AUX IN/OUT) cable 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Healthcare Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2018-RN-00241-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2018-03-26
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Siemens has identified that in some cases the ecg signal may flatline due to electromagnetic interference when used in conjunction with electrosurgical equipment. the electrosurgical equipment creates electromagnetic interference (emi) which is then transmitted to the physio module and the sc2000 system through the aux in/out cable. this interference causes the system to lose communication with the physio module. to date, there have been no reported injuries as a result of this issue.
  • 조치
    Siemens is advising: If the ECG trace disappears or flatlines, perform the following steps to disable the ECG trace and exit the physio module functionality: 1. Press F13 (Physio) on the keyboard. 2. Press the scroll wheel to select the appropriate ECG or AUX trace option and then press the Off soft key. 3. To exit the Physio menu, press F13 (Physio) again. After these steps are performed, the ultrasound system will use time-based captures instead of beat captures. Also note that the system may be rebooted to reset the ECG trace. However, if the electrosurgery equipment is used again the symptom will reoccur. This issue will be resolved by a hardware update to affected systems. Siemens Customer Service will contact users to schedule this replacement if the system configuration is affected by the scope of this issue.

Device

  • 모델명 / 제조번호(시리얼번호)
    ACUSON SC2000 ultrasound with 5.0 software (VB20A, VB20B, or VB20C) and the ECG (physio) hardware and Auxiliary (AUX IN/OUT) cableCatalogue number: 10433816ARTG Number: 139591Siemens Healthcare - Ultrasound system, imaging, cardiovascular