ACUSON X700 Ultrasound System with software versions 1.0.00, 1.0.01, 1.0.02, 1.1.00 or 1.1.02 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00420-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2014-04-08
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    When foursight 4d ultrasound imaging, 3-scape real-time 3d imaging, or anatomical m-mode are used during an obstetric or gynecology studies, the following data from the previous patient could replace the current patient's data in the report:- clinical lmp (last menstrual period)- clinical edc (estimated date of confinement)- clinical ma (menstrual age)- indicationif the replaced data is not noticed, there is a potential risk of performing additional diagnostic or non-indicated therapeutic procedures.
  • 조치
    To avoid this scenario, the following steps are recommended by Siemens: - Always review the patient report at the end of each study and confirm the report contains only measurements and patient data relevant to the exam just performed. - Before exiting fourSight 4D ultrasound imaging or 3-Scape (3D) imaging access and review the patient report. This step refreshes the data in the report. A new version of software resolving the error will be implemented in the affected units.

Device

  • 모델명 / 제조번호(시리얼번호)
    ACUSON X700 Ultrasound System with software versions 1.0.00, 1.0.01, 1.0.02, 1.1.00 or 1.1.02Catalogue/Lot number – 1065844 ARTG number: 137563
  • Manufacturer

Manufacturer