Advantx, Innova 2000, Innova 2100 IQ, Innova 3100, Innova 3100 IQ, Innova 4100, Innova 4100 IQ Cardiovascular X-ray Imaging Systems equipped with Video monitor Suspension model numbers 2236709, 2353620, 2223039-2 and 2270677-2 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 GE Healthcare Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2013-RN-00181-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2013-03-27
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    There is a potential for the locking nut to loosen out of the video monitor suspension yoke. this will not lead to an immediate fall of the suspended monitor since it will be held in place by the remaining bushing and a screw in the absence of the locking-nut.
  • 조치
    A GE representative will contact the affected customers as soon as possible to arrange to check the potentially affected systems and implement permanent corrective actions to mitigate the potential for the failure mode occurring. Meanwhile, GE is asking customers to contact their local GE Healthcare Service Representative immediately if they observe any inappropriate performance or excessive movement of the monitor suspension yoke.

Device

  • 모델명 / 제조번호(시리얼번호)
    Advantx, Innova 2000, Innova 2100 IQ, Innova 3100, Innova 3100 IQ, Innova 4100, Innova 4100 IQ Cardiovascular X-ray Imaging Systems equipped with Video monitor Suspension model numbers 2236709, 2353620, 2223039-2 and 2270677-2 ARTG number: 93871
  • 의료기기 분류등급