Advantx, Innova 2000, Innova 2100 IQ, Innova 3100, Innova 3100 IQ, Innova 4100, Innova 4100 IQ Cardiovascular X-ray imaging systems equipped with Video Monitor Suspension (Fluoroscopic, angiographic x-ray system) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 GE Healthcare Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2013-RN-00899-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2013-08-23
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    It has been identified during preventive maintenance activities that the 4 outer screws that hold the overhead video monitor suspension to the carriage were found missing since installation of the system. there is a potential that the system has not been installed per specified installation requirements. the missing screws may increase the risk for fall of the monitor suspension.
  • 조치
    GE Healthcare is inspecting all potentially affected systems to verify that the overhead Video Monitor Suspensions are properly installed and will correct all affected systems.