ADVIA 1200, 1650, 1800 and 2400 Clinical Chemistry Systems. An in vitro diagnostic medical device (IVD) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Ltd Diagnostics Division 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2013-RN-00933-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2013-09-05
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Siemens has confirmed that the affected advia chemistry systems will report results calculated using the ratio parameters feature without error flags when underlying individual test results used as part of the calculation are flagged. in cases where an error flag suppresses a numerical result, a ratio calculation will not be performed. if a numerical result is generated with a flag, the ratio will be reported without a flag.
  • 조치
    End users are requested to ensure that all ratio tests should be reviewed for flags prior to approving. Siemen's is developing a new version of software that will introduce a new flag for all ratio results that have a flag on a component test.

Device

  • 모델명 / 제조번호(시리얼번호)
    ADVIA 1200, 1650, 1800 and 2400 Clinical Chemistry Systems. An in vitro diagnostic medical device (IVD)Siemens Reference Number: CHSW 13-01ARTG Number: 174383
  • Manufacturer

Manufacturer