ADVIA Centaur CP – BNPCatalogue Numbers: 02816138 (100 tests) and 02816634 (500 tests) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Ltd Diagnostics Division 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00878-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2014-08-08
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Siemens healthcare diagnostics has received complaints of failed calibrations on the advia centaur® cp system when using bnp kit lots ending in 172. siemens has confirmed that the failed calibration is due to an error on the advia centaur cp master curve card. the master curve card error results in a low calibration deviation outside the specified calibration evaluation range when using kit lots ending in 172 on the advia centaur cp. an invalid calibration status will prevent bnp results from being generated and may delay the time to result for bnp. if the calibration is successful, the results will be valid and there is no risk to health.
  • 조치
    Siemens is advising that if a successful calibration is achieved with BNP kit lots ending in 172 and quality control results are in range, you can continue to report patient results. If unable to obtain a successful calibration on the ADVIA Centaur CP with BNP kit lots ending in 172, please switch to an alternate lot of BNP. This action has been closed-out on 31/08/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    ADVIA Centaur CP – BNPCatalogue Numbers: 02816138 (100 tests) and 02816634 (500 tests)Siemens Material Numbers (SMN): 10309044 and 10309045 Lot Numbers: lots ending in 172 (Expiration date: 17 April 2015)ARTG Number: 175075