ADVIA Centaur CP, XP, and XPT Systems with use of BNP and TSH3-Ultra 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Healthcare Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2018-RN-00033-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2018-01-18
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Siemens healthcare diagnostics has identified the following issues:- the advia centaur cp, xp and xpt systems will not expire the multi-diluent 15 pack at the end of the required 7-day on board stability (obs), if any of the affected bnp or tsh3ul lots are present on the system(s) when performing on-board dilutions; and.- on the advia centaur xpt only, the system becomes inoperable under the following scenario: 1) scanning the master curve card of advia centaur bnp kit lots ending in 193 or higher or advia centaur tsh3-ultra kit lots ending in 301 or higher, and 2) processing a bnp or tsh3ul onboard dilution of a sample. in this case, the system may display an “unknown state” message and requires customer action. when this error occurs, any tests that are in-process would have to be repeated.
  • 조치
    Siemens is advising customers who use ALL ADVIA Centaur Systems: - Do not perform onboard dilutions on any ADVIA Centaur system if any of the affected product kit lots are present on the system - Onboard dilutions for the ADVIA Centaur BNP and/or TSH3UL assay(s) are only supported when using ADVIA Centaur BNP kit lots ending in 210 and above and ADVIA Centaur TSH3UL kit lots ending in 310 and above. For customers using ADVIA Centaur XPT Systems, if users intend to perform onboard dilutions for the ADVIA Centaur BNP and/or TSH3UL assays, users must restart the workstation after scanning the master curve card for the BNP and/or TSH3UL assay kit lot, before ordering an onboard dilution for these assays. For customers using ADVIA Centaur CP Systems, if any of the affected BNP kit lots or TSH3UL kit lots have been in use on the system, users must manually track the time that the Multi-Diluent 15 has been in use on board the system.

Device

  • 모델명 / 제조번호(시리얼번호)
    ADVIA Centaur CP, XP, and XPT Systems with use of BNP and TSH3-UltraIn vitro diagnostic medical devicesADVIA Centaur BNP (100 tests)Catalogue: 02816138 SMN Number: 10309044 Kit Lot Numbers ending in: 192,193,196 ADVIA Centaur TSH3-Ultra (100 tests) Catalogue: 06491072 SMN Numbers: 10282378 Kit Lot Numbers ending in: 298, 301ADVIA Centaur TSH3-Ultra (500 tests) Catalogue: 06491080 SMN Number: 10282378 Kit Lot Numbers ending in: 298,301ARTG Number 175075
  • Manufacturer

Manufacturer