ADVIA Centaur PSA 100 test kit and 500 test kit. An in vitro diagnostic medical device (IVD) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Healthcare Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00005-2
  • 사례 위험등급
    Class III
  • 사례 시작날짜
    2016-01-06
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Siemens has confirmed that the lots ending in 263 and above are not meeting the current high-dose hook effect expectation that patient samples with total psa levels as high as 10,000 ng/ml (µg/l) will assay greater than 100 ng/ml (µg/l) as indicated in the instructions for use (ifu).Based on siemens’ internal testing, samples with total psa concentrations between 4200 ng/ml (µg/l) and 8400 ng/ml (µg/l) do not assay >100 ng/ml (µg/l) but instead result in falsely depressed concentrations of approximately 50 ng/ml (µg/l) to 94 ng/ml (µg/l).The disease state would be clinically apparent in a patient with a psa level >4200 ng/ml (ìg/l). a review of previously generated results due to this issue is at the discretion of the laboratory.
  • 조치
    The sponsor is advising users to review the customer letter. Customers can continue to use this product with the understanding that patient samples with total PSA levels between 4200 and 8400 ng/mL (µg/L) will result in falsely depressed results between 50 and 94 ng/mL (µg/L). Customers may opt to change their dilution point based on their patient population and individual laboratory needs.

Device

  • 모델명 / 제조번호(시리얼번호)
    ADVIA Centaur PSA 100 test kit and 500 test kit. An in vitro diagnostic medical device (IVD)Siemens Material Numbers: 10310292 and 10310293Lot Numbers: All lots ending in 263 and aboveARTG Number: 175243