ADVIA Centaur Systems, Dimension Vista Systems & IMMULITE Systems Estradiol Assays 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Healthcare Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00068-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2016-01-15
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Siemens healthcare diagnostics is conducting a correction for the certain estradiol products. it has been confirmed that the drug fulvestrant (faslodex) may cause falsely elevated estradiol results in the assays listed, which could lead the clinician to misinterpret the patient as pre-menopausal possibly leading to altered or discontinued use of the potential beneficial drug fulvestrant.The risk to health applies to all patients being treated with the drug fulvestrant.
  • 조치
    Siemens’ Estradiol assays may continue to be used to report results for patients not on Fulvestrant therapy; these assays should not be used for patients being treated with Fulvestrant. For patients being treated with Fulvestrant, an alternate method such as LC-MS which is not expected to show cross reactivity to Fulvestrant should be used to measure Estradiol concentrations.

Device

  • 모델명 / 제조번호(시리얼번호)
    ADVIA Centaur Systems, Dimension Vista Systems & IMMULITE Systems Estradiol AssaysADVIA Centaur Enhanced EstradiolCatalogue # and SMN: 10490889, 10491445, 10697757Dimension Vista LOCI EstradiolCat # K6463SMN: 10489099IMMULITE/ IMMULITE 1000 EstradiolCat #: LKE21, LKE21(D)SMN: 10381132, 10702832IMMULITE 2000 EstradiolCat #: L2KE22, L2KE22 (D), L2KE26, L2KE26 (D)SMN:10381178, 10702833, 10381177, 10702834All lots affectedARTG # 175698, 180109 and 179720
  • 의료기기 분류등급
  • Manufacturer

Manufacturer