ADVIA Centaur Systems Progesterone Kits 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Ltd Diagnostics Division 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00088-1
  • 사례 위험등급
    Class III
  • 사례 시작날짜
    2015-02-03
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Siemens healthcare diagnostics has confirmed a high bias on results greater than 30 ng/ml (95.4 nmol/l) when compared to the master curve when using the affected advia centaur progesterone kit lots. the high bias may cause neat serum samples with results greater than 60 ng/ml (190.8 nmol/l) to recover within the assay range upon dilution (dilution factor applied). on the advia centaur systems, for samples greater than 30 ng/ml (95.4 nmol/l), customers may observe an average dilution recovery between 48.3% and 105.0%. samples (both neat and diluted) with values below 30 ng/ml (95.4 nmol/l) do not demonstrate this bias.Progesterone values above 30 ng/ml (95.4 nmol/l) are indicative of pregnancy. the difference in neat values compared to diluted values will not change the clinical interpretation of the results. siemens advises that the severity of potential injury is negligible and the overall risk to health is negligible to non-existent. no laboratory look back is recommended.
  • 조치
    A workaround is provided to customers regarding dilution of samples for lots ending in 268 only. Technical Specialist contact number details are also provided for users as ongoing support. This action has been closed-out on 06/09/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    ADVIA Centaur Systems Progesterone KitsTest Code: PRGEKit Lot Numbers: Ending in 268Catalogue Numbers: 02382928 (1-pack), 01586287 (5-pack), 02177364 (5-pack Ref)Expiration Date: May 22, 2015ARTG Number: 175698
  • Manufacturer

Manufacturer