ADVIA Centaur XPT Systems. An in vitro diagnostic medical device (IVD) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Healthcare Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00573-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-05-03
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Siemens healthcare diagnostics has identified an issue with the advia centaur xpt default setting for the daily maintenance task (or daily cleaning procedure) frequency. instruments may have the daily maintenance task frequency set to “as needed” instead of “daily”. not performing the daily maintenance may impact any assay and can be detected through monitoring of quality control and calibration results.
  • 조치
    Siemens is advising users to not release results generated by the system if quality control results are not within acceptable ranges, and is providing instructions for users to check the daily maintenance tasks. If the frequency is set to "As Needed", users should contact Siemens in order for the frequency to be changed. All systems will be checked during the next scheduled maintenance. Siemens is not recommending a review of previously generated results. However, a review of previously generated results is at the discretion of the Laboratory Director.

Device

  • 모델명 / 제조번호(시리얼번호)
    ADVIA Centaur XPT Systems. An in vitro diagnostic medical device (IVD)Siemens Material Number (SMN): 10711433ARTG Number: 175890
  • Manufacturer

Manufacturer