ADVIA Chemistry Triglyceride Concentrated (TRIG_c) Reagent. An in vitro diagnostic medical device (IVD) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Healthcare Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-01103-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-08-22
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Siemens healthcare diagnostics has confirmed the advia chemistry triglyceride concentrated reagent (trig_c) does not meet instructions for use (ifu) linearity claim at the upper limit of the assay range as it approaches the end of its shelf life. internal testing indicates that a patient bias (-7% to -12%) may be seen at concentration levels from 6.22 to 6.78 mmol/l. linearity is reduced by approximately 50% at triglyceride concentrations approaching the upper end of the extended range of 12.43 mmol/l.Quality control (qc) material may not detect this issue depending on the triglyceride concentrations in the qc material.Siemens internal testing has confirmed all other in date lots are currently meeting linearity ifu claims. however, as a precaution, until root cause is determined, a short term mitigation of reducing the expiration date to 6 months for all lots of advia chemistry systems triglyceride concentrated has been implemented.
  • 조치
    Siemens is advising users to discontinue use and discard any remaining stock of lots 365096 and 379597. Discontinue use and discard remaining lots once they reach the revised shelf life. Customers will need to manually monitor the revised expiration dates for the impacted lots. A review of previously generated results is at the discretion on the Laboratory Director.