Affinis Inverse inlay ceramys & Affinis Fracture ceramic heads (Total Shoulder System that is designed for treating shoulder conditions) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Mathys Orthopaedics Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-01305-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2014-12-22
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The manufacturer, mathys ltd bettlach, received reports of disconnection of affinis inverse inlay ceramys and affinis fracture ceramic heads. subsequent investigations have identified that the inlay and individual implant components complied with the defined specifications.For affinis fracture ceramic heads:the disconnections may have resulted from inadequate primary fixation due to insufficient head-cone connection or by protruding bone residue, cement residue or soft tissue fragments inside the cone connection.For affinis inverse inlay ceramys: the disconnection incident was caused by insufficient fixation of the inlay on the cone due to lack of impaction. the affinis inverse inlay ceramys has not been supplied in australia.To minimise occurrences of disconnection, additional instructions and notes regarding ceramic inlay implantation and ceramic head implantation have been added to the surgical technique and the instructions for use (ifu).
  • 조치
    Mathys Ltd Bettlach is requesting the customers to replace all Surgical Techniques and Instructions for Use (IFUs) for Affinis Inverse and/or Affinis Fracture with the updated versions provided. The sponsor has directly contacted the surgeon to notify of the implications of inadequate fixation and to provide the updated instructions. Further information will be published on the TGA web site at http://www.tga.gov.au/current-year-alerts This action has been closed-out on 14/07/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Affinis Inverse inlay ceramys & Affinis Fracture ceramic heads (Total Shoulder System that is designed for treating shoulder conditions)Affinis Inverse inlay ceramysItem numbers: 62.34.0066, 62.34.0067, 62.34.0068, 62.34.0069, 62.34.0070 62.34.0071, 62.34.0072, 62.34.0073 & 62.34.0074 Affinis Fracture ceramic headsItem numbers: 60.25.0042, 60.25.0045 & 60.25.0048 All Batch numbersARTG number: 215501