Affixus Hip Fracture Nail 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Zimmer Biomet Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00559-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-05-01
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Zimmer biomet has identified an issue with the affixus hip fracture nail not properly engaging with the targeting jig. an investigation into the nail/targeting jig assembly showed that certain specifications were undersized causing interference to the fit of the nail onto the targeting jig which can result in a surgical delay.
  • 조치
    Users are asked to:- 1. Review the notification and ensure affected personnel are aware of the contents; 2. Quarantine all affected product for collection by your Zimmer Biomet sales representative; 3. Complete and return the acknowledgement form as per the instructions in the customer letter; and 4. Notify Zimmer Biomet of any adverse events associated with this device.

Device

  • 모델명 / 제조번호(시리얼번호)
    Affixus Hip Fracture NailMultiple Item NumbersARTG Number: 212235
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 의견
    “The safety of patients has always been, and continues to be, Zimmer Biomet’s top priority and it is our honor to be a leader in this industry for the past 90 years,” Zimmer Biomet told ICIJ in a statement. “We adhere to strict regulatory standards, and work closely with the FDA and all applicable regulatory agencies in each of our regions as part of our commitment to operating a first-rate quality management system across our global manufacturing network. The company added that it is focused on staying at the forefront of innovation and doing right by the millions of patients who rely on the company’s products.
  • Source
    DHTGA