Alaris Syringe Pumps (GH, CC, TIVA, PK and Enteral) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 CareFusion Australia 316 Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00049-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2017-04-27
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Based on overseas reports from a customer, bd/carefusion identified regarding a potential risk of syringe siphonage with pumps that have a broken “plunger back plate spring” in the plunger back plate assembly. a breakage of the plunger backplate spring may allow movement of the syringe plunger within the plunger holder mechanism which could result in siphonage. in some circumstances, this may result in a clinically significant over infusion especially in neonatal and paediatric patients, or those receiving critical drugs, at low infusion rates.
  • 조치
    CareFusion is advising users to check the age of the pump by referring to the label on the rear case and to replace the plunger backplate spring (refer to Information Notice IN0221) on syringe pumps older than three years. Clinical areas such as neonatal, paediatric and critical care, where critical drugs are delivered at lower infusion rates are to be prioritised. In the event that users see a "Check Syringe" alarm without an identifiable cause, they are requested to remove the pump from clinical use for examination by a qualified service personnel in accordance with the Alaris Syringe Pump Technical Service Manual.

Device

  • 모델명 / 제조번호(시리얼번호)
    Alaris Syringe Pumps (GH, CC, TIVA, PK and Enteral) - All VariantsProduct Codes: with prefix 8001, 8002, 8003, 8004 and 8005 (all variants)All Serial NumbersARTG Number: 177501
  • Manufacturer

Manufacturer