Alaris System PC Unit (PCU) model 8015, software version 9.12 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 CareFusion Australia 316 Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2013-RN-00707-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2013-07-12
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    In the affected devices, the keyboard processor is operating at an incorrect voltage. the correct voltage should be at 1.8 volts however the affected units are at 1.4 volts. due to this issue the alaris system pcu software v9.12 while in operation could potentially experience a loss of communication between the pcu main processor and keyboard processor. the alaris system pcu will display either a system error (error codes 111.2000 or 111.2002) or channel disconnected error message. along with the pcu display the alaris system pcu also produces an audible alarm and the attached modules will display a scrolling "communication error" message. during the communication error, the pump will alarm both visually and audibly.
  • 조치
    Customers are asked to weigh the risk/benefit to patients before continuing to use the affected PCU, version 9.12. Either discontinue use of affected PC units until remediated or continue use on patients by implementing work around instruction identified in the customer letter. CareFusion Technical Service will provide a software update to correct the voltage on the keyboard processor for those affected Alaris System PC units.

Device

  • 모델명 / 제조번호(시리얼번호)
    Alaris System PC Unit (PCU) model 8015, software version 9.12Multiple serial numbers affectedARTG Number: 146664