Alaris System Pump module (LVP) 8100 and Air-In-Line (AIL) Sensor Kits Alaris System Pump module (LVP) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 CareFusion Australia 316 Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00140-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-01-30
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The manufacturer has received reports of air-in-line (ail) alarms that have occurred when no air is observed in the line. while the pump is infusing, the system may indicate that an ail alarm has occurred when no air is in the line. in some cases, these false ail alarms may be attributed to a faulty ail sensor.To date, bd is not aware of any report of injury attributed to this defect in australia.
  • 조치
    If AIL alarm occurs, the user should do the following: 1. Determine if there is air visible in the tubing that has caused the alarm. If there is air visible in the line, clinician can evaluate whether or not it is clinically significant and remove it according to hospital protocol, if necessary. 2. If no air is observed, ensure that tubing is properly installed in AIL Detector. False AIL alarms may occur if tubing is not properly installed. When inserting the tubing into the AIL detector, use a fingertip and firmly push the tubing toward the back of the AIL detector. 3. If AIL alarms continue to reoccur on the same pump, after air has been removed from the line and tubing has been properly loaded, the AIL sensor may be faulty. If the AIL sensor needs to be replaced, BD will provide replacement parts at no charge. If a pump has been identified as having recurring AIL alarms without evidence of air in the line, then replace the AIL sensor with a new AIL sensor provided by BD.

Device

  • 모델명 / 제조번호(시리얼번호)
    Alaris System Pump module (LVP) 8100 and Air-In-Line (AIL) Sensor Kits Alaris System Pump module (LVP) Model: 8100Manufactured between October 2011 and June 2015ARTG Number: 146664Air-In-Line (AIL) Sensor Kits Part Numbers: 147083-102, 49000221 Distributed between October 2011 and June 2015ARTG Number: 146664