AlignRT patient positioning/tracking video camera system (when used withTrueBeam Radiotherapy Delivery System) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 alphaXRT Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00936-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2016-09-30
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Vision rt has been notified that unintended changes can be made to the planned couch rotation parameter during patient set-up. this can occur when a user sets a different couch rotation from the plan couch rotation while performing patient alignment with the alignrt device prior to treatment with truebeam radiotherapy delivery system.When a request to update the couch position is sent to the truebeam via the alignrt device using the move couch function, the request may not be processed as expected. if the couch rotation in the alignrt device does not match the planned couch rotation of the field selected on the truebeam, this can result in a significant change to the planned couch rotation value. when the planned value is changed, the original plan value is overwritten at the console. this will result in the treatment plan couch value being updated unintentionally and a treatment misadministration may occur. there have been no reports of patient harm due to this issue.
  • 조치
    Vision RT is advising users to not use the “Move Couch” function if the couch rotation value set in the AlignRT device does not match the planned couch rotation for the treatment field. Vision RT is investigating a solution for this issue.

Device

  • 모델명 / 제조번호(시리얼번호)
    AlignRT patient positioning/tracking video camera system (when used withTrueBeam Radiotherapy Delivery System)AlignRT PCR Number: 249-1526ARTG Number: 159239
  • 의료기기 분류등급