Alkaline Phosphatase (ALPAMP, ALPDEA and ALPA_c) used on the ADVIA 1200, 1650, 1800, 2400, and XPT Chemistry Systems Alkaline Phosphatase ALPAMP 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Healthcare Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00961-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2015-09-25
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The reagent kit lots listed above demonstrate an increased incidence of absorbance flags (u, u) and //// errors. absorbance flags u and u indicate abnormal high absorbance exceeding either the blank or sample limit. absorbance flag //// indicates a calculation error and no result is generated by the system.The increased rate of absorbance flags with these lots has not been associated with inaccurate results. if calibration and quality control results are acceptable, reported patient results are acceptable. this risk to health is negligible. siemens advises the potential for clinical impact and overall risk to health as a result of this issue is negligible.
  • 조치
    Customers are asked to quarantine affected units from use prior to destruction and/or disposal.

Device

  • 모델명 / 제조번호(시리얼번호)
    Alkaline Phosphatase (ALPAMP, ALPDEA and ALPA_c) used on the ADVIA 1200, 1650, 1800, 2400, and XPT Chemistry Systems Alkaline Phosphatase ALPAMPCatalogue Number: 03035814Lot Number: 327460Alkaline Phosphatase ALPDEACatalogue Number: 03036535Lot Number: 327465 Alkaline Phosphatase Concentrated ALPA_cCatalogue Number: 06860450Lot Number: 327471ARTG Number: 176331