All Carestation 620, 650 and 650c Anaesthesia systems 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 GE Healthcare Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-01300-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2017-10-11
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Ge healthcare has recently become aware that an incomplete seal can exist between the co2 absorber and the breathing circuit co2 bypass port assembly of the carestation 600 series systems. an incomplete seal can allow rebreathing of patient gases that have bypassed the co2 absorbent material and could result in unintended elevated levels of inspired co2 (fico2), which could lead to hypercarbia. to date, there have been no injuries reported as a result of this issue.
  • 조치
    GE Healthcare is advising users as an interim solution to increase the flow of fresh gas to reduce the volume of patient gas that is rebreathes if elevated FiCO2 levels are observed. If the FiCO2 levels cannot be adequately reduced with this action, users should consider switching to another anaesthesia delivery device. GE is releasing revised parts that minimise the likelihood of incomplete gas flow through the CO2 absorbent canister. GE will be in contact with affected customers to arrange for correction. Going forward, the updated parts will be included in the 12-month preventative maintenance replacement schedule for the Carestation 600 Series system.