All GemStar Family of Infusers: GemStar Infusion System - Single Therapy, GemStar 7 Therapy Infusion System 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Hospira Pty Limited 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2013-RN-00245-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2013-04-11
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
    ICU Medical, Inc. acquired from Pfizer in early 2017, Hospira Infusion Systems, the portion of Hospira dedicated to develop infusion pumps.
  • 데이터 추가 비고
  • 원인
    Leakage from the aa battery causes a barrier to the power contacts resulting in no power being delivered, and in some instances the leakage results in contamination or corrosion of the power circuitry. excessive voltage ingress (voltage from the outside ac power adapter) to voltage regulators can cause them to fail. failure of these regulators may result in the device shutting off without issuing a warning or an audible or visual alarm. if the device shuts off it will result in a delay/interruption in therapy.
  • 조치
    As directed by the Gemstar Technical Service Manual (TSM), the internal AA batteries and battery compartment should be inspected for signs of leakage, corrosion or other damage prior to each use. In addition, each time the batteries are replaced the battery compartment should be inspected for damage. If a device exhibits damage caused by leaking batteries, immediately remove it from clinical service and return it to Hospira for repair. Healthcare professionals are advised to weigh the risk/benefit to patients associated with the use of the device when administering critical therapies. Customers should consider the use of an alternative pump, particularly in patients in which delay/interruption in therapy or under/over infusion could result in significant injury or death.

Device

  • 모델명 / 제조번호(시리얼번호)
    All GemStar Family of Infusers: GemStar Infusion System - Single Therapy, GemStar 7 Therapy Infusion SystemGemStar Infusion System - Single TherapyCatalogue number: 611308801Number on pump: 13088/13150GemStar 7 Therapy Infusion SystemCatalogue number: 611308701Number on pump: 13087/13000