All Magnetom MRI systems with superconductive magnets 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Healthcare Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-01251-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2015-12-22
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    In some cases the quench lines of some superconducting magnets may not have been properly installed.Quench lines are installed to vent helium gas in the event of a quench (i.E., sudden loss of superconductivity when the temperature is raised). in the unlikely event that a quench were to occur, a damaged/inappropriately installed quench line could lead to a displacement of oxygen, cold burns and/or could pose a risk to users, patients or others if the helium gas is vented to a closed area inside a building.
  • 조치
    The sponsor is advising users to contact their contract quench line installer (Siemens or third party) to perform a quench line inspection. In the interim, users are reminded that it is important to follow the guidelines outlined in the System Owner Manual, including ensuring that they have an emergency plan in place in the event helium gas escapes into the magnet room or other rooms. This action has been closed-out on 04/05/2017.

Device

  • 모델명 / 제조번호(시리얼번호)
    All Magnetom MRI systems with superconductive magnetsVarious System NamesCatalogue Numbers: 10433372, 10432914, 7391167, 10849579, 10018165, 10352127, 10836666, 10836618, 10849582, 10849583, 10432915, 10849580, 7104719, 10655588, 7104594, 106557, 10018223, 7387074, 10018221 and 10276755ARTG Numbers: 98319, 144221, 98485, 154128 and 188470
  • Manufacturer

Manufacturer