All Plum A+ Family of Infusers: Plum A+Single Channel Infusion Pump, Plum A+ and A+3 Infusion Pumps with MedNet Software 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Hospira Pty Limited 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2013-RN-00188-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2013-02-27
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
    ICU Medical, Inc. acquired from Pfizer in early 2017, Hospira Infusion Systems, the portion of Hospira dedicated to develop infusion pumps.
  • 데이터 추가 비고
  • 원인
    The distal pressure sensor calibration may have drifted and require re-calibration. the following error codes may be displayed during set up or infusion on the affected devices: e180/n180, e181/n181, e186/n186, e187/n187, e346. if distal pressure sensor drift occurs, the pump may not sense the build up of pressure and will not alarm when occlusion thresholds are exceeded.
  • 조치
    Hospira is providing end users with instructions to determine if they are affected by the issue. Healthcare professionals are advised to weigh the risk/benefit to patients associated with the use of the device when administering critical therapies. Hospira is also in the process of evaluating the design of the distal pressure sensor to reduce the opportunity for pressure sensor drift. Additionally, a yearly test to determine if the distal pressure sensor’s calibration has drifted will be released in 2013.

Device

  • 모델명 / 제조번호(시리얼번호)
    All Plum A+ Family of Infusers: Plum A+Single Channel Infusion Pump, Plum A+ and A+3 Infusion Pumps with MedNet SoftwarePlum A+Single Channel Infusion PumpCatalogue Number: 611239101Number on Pump: 12391 & 11971Plum A+ Infusion Pump with MedNet SoftwareCatalogue Number: 612079201Number on Pump: 20792Plum A+3 Infusion Pump with MedNet SoftwareCatalogue Number: 612067801Number on Pump: 20678