Allen Wrench 5mm in SMR supplied in instrument set (surgical instrument used during implantation of SMR anatomic and Reverse shoulder prosthesis) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Lima Orthopaedics Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-01053-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-08-10
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The manufacturer, limacorporate has received two complaints about the allen wrench. firstly, the impossibility to connect the allen wrench to the t-handle intra-operatively. secondly, after a difficult insertion, it was impossible to remove the allen wrench from the t-handle. in both cases, there were no reported consequences for the patients. limacorporate's investigations have shown that the diameter of the zimmer connection on the allen wrench has been manufactured out of specification and is over-dimensioned. this leads to difficult/impossible coupling between the allen wrench and the t-handles.
  • 조치
    Lima Orthopaedics' kit coordinator will be visiting hospitals and replacing affected devices in the SMR instrument sets with unaffected devices. This action has been closed-out on 20/02/2017.