Allura Systems: FD10, FD20, FD10/10, Allura Systems with OR table: FD10, FD10/10, FD20/20 (Fluoroscopic x-ray systems used for diagnostic/interventional procedures in cardiovascular and vascular imaging applications) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Philips Electronics Australia Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2013-RN-00588-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2013-06-12
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Philips has received reports of occurrences where the monitor ceiling suspension (mcs) system fell to its lowest position due to an assembly error of the mcs actuator. in one occasion it collided with the table top.
  • 조치
    Philips Healthcare is issuing work around instructions as a temporary measure. The affected MCS actuators will be replaced as a permenant fix.

Device

  • 모델명 / 제조번호(시리얼번호)
    Allura Systems: FD10, FD20, FD10/10, Allura Systems with OR table: FD10, FD10/10, FD20/20 (Fluoroscopic x-ray systems used for diagnostic/interventional procedures in cardiovascular and vascular imaging applications)Product codes: 722010, 722011, 722012, 722013, 722014, 722015, 722019, 722020, 722022, 722023, 722024, 722025Applies to:Actuator Short Assy: 989600-184-665Actuator Long Assy: 989600-184-654ARTG number: 98560, 175708
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • Source
    DHTGA