Allura Xper and Allura Clarity - software releases 1.2.7, or 2.0.6, or 7.2.x where x<8 or 7.6.x or 7.8.x or 8.1.x where x< 16 or 8.2.x where x<16 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Philips Electronics Australia Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-01104-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2015-11-13
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    It has been discovered through customer complaints and internal testing that there is an intermittent electronic product defect. in certain circumstances, a software error can lead to a situation where the five minute fluoroscopy audible signal does not sound. no injuries attributed to the problem have been reported.
  • 조치
    The sponsor will perform a software update to the affected devices. In the meantime, the user is advised to always observe real time dose information and cumulative fluoro time provided by the system. The fault condition is reset when a new patient case is started or when the system is restarted. This action has been closed-out on 05/12/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Allura Xper and Allura Clarity - software releases 1.2.7, or 2.0.6, or 7.2.x where x<8 or 7.6.x or 7.8.x or 8.1.x where x< 16 or 8.2.x where x<16Product Codes: 722026, 722027, 722028, 722029, 722038, 722058, 722033, 722034, 722035, 722036, 722039, 722059, 722003, 722005, 722006, 722008, 722010, 722012, 722013, 722015, 722023.Allura Xper FD R2.xProduct Code: 722006Allura Xper FD R8.2.0 Product Code: 722035ARTG Number: 175708
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • Source
    DHTGA