Allura Xper with FlexVision large screen monitor 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Philips Electronics Australia Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2018-RN-00830-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2018-06-25
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Philips has received reports of instances where after a continuous operation of more than one and a half days, the image on the flexvision large screen monitor froze for about 15 seconds. thereafter, the system restored itself and was fully operational again.This is only applicable to allura systems with a flexvision large screen monitor. all other monitors to the system are not affected.To date, philips is not aware of any injuries that may have occurred due to this issue.
  • 조치
    Philips is advising this issue will be resolved in a software upgrade, expected to be available by the second half 2018. Users will be contacted by Philips when the software is available for installation. In the interim, users are advised to restart the system at least once per day to prevent occurrence of this issue.

Device

  • 모델명 / 제조번호(시리얼번호)
    Allura Xper with FlexVision large screen monitorSystem names: Allura 8.1.25.0, Allura 8.1.25.1, Allura 8.1.25.5, Allura 8.2.25.0, Allura 8.2.25.5 and Allura 8.2.27 ARTG Number: 225815(Philips Electronics Australia Ltd - X-ray system, diagnostic, fluoroscopic, angiographic, stationary, digital)
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • Source
    DHTGA