ALT and Glucose-Stat Assays (used for the determination of ALT and Glucose). An in vitro diagnostic medical device (IVD) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Beckman Coulter Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00758-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-06-10
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Previously unknown drug interferences from sulfasalazine and sulfapyridine have been identified. there is a possibility that drug interference's from administration of sulfasalazine to patients may generate false low alt results and false low glucose-stat results. sulfapyridine may generate false high glucose-stat results (osr6x40).Interference seen with 300mg/l sulfasalazine showed a negative bias of up to 35% for the low alt control and up to 29% for the high alt control.Interference seen with 300mg/l sulfasalazine showed a negative bias of up to 25% (or 1.3 mmol/l) for the low glucose-stat control and up to 13% (or 1.6mmol/l) for the high glucose-stat control.Interference seen with 300mg/l sulfapyridine showed a positive bias of up to 21% (or 1.1 mmol/l )for the low glucose-stat control and up to 9% or (1.1mmol/l) for the high glucose-stat control.There is no risk to patients however there is the potential for a delay in results.
  • 조치
    Beckman Coulter is advising users that the IFU has been updated to include additional warnings in the 'Interfering Substances' section. No retrospective review of results is recommended.

Device

  • 모델명 / 제조번호(시리얼번호)
    ALT and Glucose-Stat Assays (used for the determination of ALT and Glucose). An in vitro diagnostic medical device (IVD)Reference numbers: ALT OSR6X07, Glucose-Stat OSR6X40All lots affectedARTG numbers: 200015, 200018