Application Instrument for Sternal ZIPFIX(Sternal ZIPFIX is used for closure of the sternum following sternotomy to stabilize the sternum and promote fusion) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Johnson & Johnson Medical Pty Ltd t/a DePuy Synthes 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00897-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2015-09-15
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    In the affected lots of the application instrument for sternal zipfix :· the end cap may loosen, thus reducing the tension applied to the implant. · the end cap may detach, allowing the tensioning spring to also become detached, making the instrument non-functional.If the end cap is loose, the maximum tension applied to the implant is reduced and may lead to insufficient sternal bone reduction. if the tension coil spring detaches completely from the zipfix application instrument while closing the sternum, it is possible that the spring or nut could fall into the thoracic cavity and go undetected. if the nut/spring is retained in the thoracic cavity, adverse tissue reaction may occur. no such occurrence has been reported to date.
  • 조치
    Customers are requested to immediately check their inventory to determine if the facility has any affected product and if so to quarantine those units prior to returning them to JJM for repair. This action has been closed-out on 05/09/2016.