AQUIOS CL Flow Cytometry System. An in vitro diagnostic medical device (IVD) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Beckman Coulter Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-01101-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-08-23
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    As a result of internal investigations, beckman coulter has learned that an erroneous configuration setting during servicing may be possible on the aquios cl instruments. if the incorrect configuration occurs, the probe may aspirate samples only from a single tube and will not aspirate samples from the remaining tubes in the cassette. the probe would repeatedly aspirate from a single tube position in the cassette, resulting in erroneous patient results for the remaining tubes in the cassette. this causes a mismatch between the test result given and the sample id.
  • 조치
    1. Quality Control samples must be run utilising the control cassette and the autoloader. The controls will not pass if this issue exists on the unit, when run using the autoloader. If the controls do not pass, immediately contact a Beckman Coulter representative. 2. BEC Service organisation will be in contact to ensure that the configuration settings are correct.

Device

  • 모델명 / 제조번호(시리얼번호)
    AQUIOS CL Flow Cytometry System. An in vitro diagnostic medical device (IVD)Part Number: B30166ARTG Number: 177999