AQUIOS CL Flow Cytometry System. An in vitro diagnostic medical device (IVD) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Beckman Coulter Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-01325-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-10-16
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Beckman coulter has determined that when aquios cl systems are connected to a laboratory information system (lis), the system may duplicate sample requests leading to sample mis- identification. for this to occur, the following conditions must be met:- the aquios cl flow cytometer is connected to an lis; and- the aquios cl system has the default test enabled; and- host query is enabled in lis setup screen on system; and- there are multiple cassettes in the autoloader; and- the lis host query response is received in a brief time window before lis response timeout occurs.When all of these conditions are met, a sample will follow two paths of test request creation (lis and default test). all software versions (2.0, 2.0.1 and 2.1) in existence are impacted by this issue and all applications used on the aquios cl are impacted by this issue.
  • 조치
    Beckman Coulter is advising users to immediately perform an adjustment to the software options on the system by turning off the DEFAULT TEST. This will eliminate the issue and result in one sample producing one set of results. After doing this, samples with no LIS test requests will become entries in the incomplete tab. Beckman Coulter will be contacting laboratories to assist in determining if any samples previously run are impacted.

Device

  • 모델명 / 제조번호(시리얼번호)
    AQUIOS CL Flow Cytometry System. An in vitro diagnostic medical device (IVD)Part Number: B30166All software versions (2.0, 2.0.1 and 2.1)All applications used on the AQUIOS CL SystemARTG Number: 177999
  • Manufacturer

Manufacturer