AQURE Data Management System Software prior to version 1.8.2 may be affected. 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Radiometer Pacific Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00453-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2014-04-17
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Radiometer has recently become aware of three potential issues with unit conversion in aqure that may occur if the device is transmitting in a different unit than aqure is displaying and/or transmitting to the his/lis. - conversion of ketone from mmol/l to mg/dl or vice versa: the parameter ketone is measured on either the siemens clinitek or the docureader. this is not applicable in australia as none of these devices are connected to aqure. - conversion of glucose and lactate from g/l to mmol/l or vice versa: in both cases a potential wrong conversion factor may be used, which may cause the reported result to be incorrect by a factor 10,000. this problem may be present in all versions of aqure.- converting from mg/dl to mg/l and g/dl to g/l or vice versa: a potential wrong conversion factor may be used of 1/10. the correct factor is 10. reported results may be incorrect by a factor 100. this problem is not parameter specific and may be present in all versions of aqure.
  • 조치
    End users are requested to not change the units on AQURE or iSTAT without consulting Radiometer Pacific. Radiometer will be carrying out testing to identify affected software and a software update will be installed to correct the issue. This action has been closed-out on 29/01/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    AQURE Data Management SystemSoftware prior to version 1.8.2 may be affected.