ARCHITECT DHEA-S Reagents. An in vitro diagnostic medical device (IVD) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Abbott Australasia Pty Ltd Diagnostic Division 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-01526-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-12-18
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Abbott has identified that falsely elevated results may be obtained when using the architect dhea-s assay with samples from infants up to 60 days old. the specific cause of the elevated results is currently under investigation. although falsely elevated results have only been observed with samples from infants up to 60 days old, the expected values representative data for children up to 10 years old provided in the architect dhea-s reagent insert should not be used as these reference ranges are being reviewed as part of the investigation.All in-date lots of current reagent inventory are impacted.
  • 조치
    Abbot is advising users to immediately discontinue use of the ARCHITECT DHEA-S assay with samples from infants up to 60 days old and to immediately discontinue use of the ARCHITECT DHEA-S reagent insert Expected Values representative data for children up to 10 years old. Users are advised to review this letter with the Medical Director and follow laboratory protocol regarding the need for reviewing previously reported patient results.

Device

  • 모델명 / 제조번호(시리얼번호)
    ARCHITECT DHEA-S Reagents. An in vitro diagnostic medical device (IVD)List Number: 8K27-20 Lot Numbers: 01316K000, 01217A000, 02217D000, 04117H000List Number: 8K27-25Lot Numbers: 01216K000, 01117A000, 02417D000, 00517F000ARTG Number: 196075
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 의견
    “We are in constant communication with regulatory agencies and competent authorities worldwide which allows us to implement global recalls or in-country communication quickly and effectively,” Abbott, which now owns St. Jude Medical told ICIJ in a statement. In addition to sending global notices to physicians worldwide, we also make sure that product advisories are available online and classification of product recalls and product advisories are determined by global regulatory bodies which can impact the timing in any given country. MD companies follow varying regulations in different countries. In come countries software is not regulated so a recall in one country related to software would not be classified as a recall or field action in another. In addition, review cycles within the regulatory process can be different in each country which can impact communication and recall timing.