Architect Sex Hormone Binding Globulin (SHBG) reagents. An in vitro diagnostic medical device (IVD) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Abbott Australasia Pty Ltd Diagnostic Division 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00519-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-04-26
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Abbott has identified that architect shbg results generated with the potassium edta specimen tube type may demonstrate a negative shift relative to results generated with serum. shbg dimer destabilisation in edta could result in low shbg measurements by immunoassay. in other words, there is a potential for falsely decreased architect shbg results to be generated when using the architect shbg assay with the potassium edta specimen tube type. internal studies have determined the use of the potassium edta specimen tube type may result in a decrease in shbg concentration values of greater than 20% when compared with serum collected in serum tubes. the shift was observed across the full analytical range of the assay. as a result, an increase in free androgen index / free testosterone index calculations of up to 30% may also be observed when using potassium edta specimen tubes.
  • 조치
    Abbott Diagnostics is advising users to immediately discontinue the use of the Potassium EDTA specimen tube type with the ARCHITECT SHBG assay. Users are requested to review the recall notification with their Medical Director to ensure that they are aware of this issue and to determine if a review of previously generated results using the Potassium EDTA specimen tube type is required.

Device

  • 모델명 / 제조번호(시리얼번호)
    Architect Sex Hormone Binding Globulin (SHBG) reagents. An in vitro diagnostic medical device (IVD)List Number: 8K26-20Lot Numbers: 01916E000, 00316F000, 01816G000, 00916I000, 07316I000 and 00916L000List Number: 8K26-25Lot Numbers: 01816E000, 00216F000, 01716G000, 00816I000 and 07216I000ARTG Number: 196075
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 의견
    “We are in constant communication with regulatory agencies and competent authorities worldwide which allows us to implement global recalls or in-country communication quickly and effectively,” Abbott, which now owns St. Jude Medical told ICIJ in a statement. In addition to sending global notices to physicians worldwide, we also make sure that product advisories are available online and classification of product recalls and product advisories are determined by global regulatory bodies which can impact the timing in any given country. MD companies follow varying regulations in different countries. In come countries software is not regulated so a recall in one country related to software would not be classified as a recall or field action in another. In addition, review cycles within the regulatory process can be different in each country which can impact communication and recall timing.
  • Source
    DHTGA